In a children’s insulin pump trial, only 2 of 31 high morning ketone readings followed a glucose high that would prompt a check
Why it matters
In a trial where children using insulin pumps checked ketones every morning, only two of 31 elevated results followed the long or extreme sugar high that would prompt a check in two later trials. To compare pumps, it matters when families were asked to test: a day without a ketone test can’t be counted as a day with normal ketones. These results do not tell us which pump better prevents a medical emergency.
Insulin pump trials count days with high blood ketones to help assess safety. Ketones form when the body uses fat for fuel. In type 1 diabetes, they can rise during fasting or when a pump stops delivering insulin. To compare counts across trials, it matters when families test: some trials ask for a ketone test every morning, while others ask families to check after a prolonged or extreme glucose rise. How often would the overnight glucose readings have prompted a check on the mornings when scheduled tests found high ketones?
Across 2,594 monitored nights in a pediatric trial, 31 morning blood tests showed ketones above 1.0 millimoles per liter (mmol/L). On only 2 of those nights did a glucose sensor record a high that met the rule for a check in two later pump trials. On 23 of the 31 nights, glucose never exceeded 300 milligrams per deciliter (mg/dL) on the sensor.
When families tested
In the pediatric pump shut-off trial (PSO4), children aged 4–14 used a pump that could pause insulin delivery when a sensor predicted low glucose. Each night, the trial randomly enabled or disabled the automatic pause. Families were asked to test blood and urine ketones before breakfast after either kind of night. They were also asked to test blood ketones whenever a fingerstick blood glucose measurement reached 250 mg/dL. Wadwa and colleagues found more morning ketone elevations in younger children and examined their relation to waking glucose. A waking measurement alone cannot show whether glucose stayed high for hours and then fell before breakfast.
Two later trials of automated insulin delivery, DCLP3 in adolescents and adults and DCLP5 in children, specified a two-step check. If sensor glucose stayed above 300 mg/dL for more than two hours or reached 400 mg/dL once, the family was to check glucose by fingerstick. If the fingerstick exceeded 300 mg/dL, the family was to test blood ketones (trial protocols, DCLP3 p. 40 and DCLP5 pp. 45–46).
We asked which overnight glucose records in PSO4 met the first step of that rule and compared them with the next morning’s ketone results. We counted blood ketones strictly above 1.0 mmol/L as elevated because PSO4 tracked that threshold and DCLP3 reported days above it.
Overnight glucose and morning ketones
Of 3,606 randomized nights in PSO4, 3,500 had a morning blood ketone result and 2,594 also had sufficient sensor readings to assess overnight glucose. These nights included 31 elevated morning results in 19 of 82 children. Bedtime glucose had to be 90–270 mg/dL to start monitoring. A sensor alarm at 250 mg/dL and instructions to test ketones at a fingerstick of 250 mg/dL or higher gave families a chance to act before a high became prolonged.
Only 44 of the 2,594 nights (1.7%) met the later trials’ sensor rule (Table 1). An elevated morning ketone result was more frequent after a qualifying night (2/44, 4.5%) than after other nights (29/2,550, 1.1%). But qualifying nights were so rare that they accounted for just 2 of the 31 elevations (6.5%). The approximate 95% interval, estimated by resampling children, was 0–19%: the exact fraction is uncertain, but even the upper estimate represents a minority of elevated mornings. One of 16 elevated mornings with automatic pauses enabled followed a qualifying night, as did one of 15 with pauses disabled.
Table 1. Scheduled morning blood tests by the preceding night’s sensor glucose. Each night appears in one glucose category. The last column counts a subset of the elevations in the preceding column.
| Overnight sensor glucose (mg/dL) | All nights | Blood ketones >1.0 mmol/L | Blood ketones ≥1.5 mmol/L |
|---|---|---|---|
| Peak ≤250 | 1,830 | 15 | 2 |
| Peak >250 to 300 | 547 | 8 | 4 |
| Peak >300, neither prolonged nor extreme | 173 | 6 | 3 |
| Above 300 for >2 hours or peak ≥400 | 44 | 2 | 1 |
| Total | 2,594 | 31 | 10 |
Six elevated mornings followed a rise above 300 mg/dL that was too brief to qualify. On all six nights, glucose first crossed 300 more than two hours before monitoring ended, so a late start alone cannot explain why the rises did not qualify. The longest uninterrupted stretches above 300 lasted 10–95 minutes, compared with 135 and 185 minutes on the two qualifying nights with elevated morning ketones. Even a rule calling for a check after a single sensor reading above 250 would have selected 764 nights and included only 16 of the 31 elevations.
Both waking glucose and overnight peaks tended to be higher on mornings with elevated ketones than on mornings without them. Median waking glucose was 246 versus 153 mg/dL, and the median overnight peak was 254 versus 216 mg/dL. Yet waking glucose varied even among the 29 elevated mornings without a qualifying overnight high. On six it was 43–98 mg/dL, consistent with ketone production during fasting. On six others it was above 250 but no higher than 300 mg/dL.
Fingersticks gave families another way to detect high glucose. On nine of the 29 elevated mornings without a qualifying sensor high, families had a fingerstick above 300 mg/dL during the night or at waking. Of the two qualifying nights with elevated morning ketones, one had a confirming fingerstick near the qualifying sensor reading; the other had no nearby meter reading.
Morning ketone elevations were most frequent in the youngest children. In the trial roster’s 3–6 age group, 8 of 14 children accounted for 17 elevated mornings in 385 nights (4.4%). At ages 7–10, 6 of 23 children had elevations on 8 of 781 nights (1.0%); at ages 11–14, 5 of 45 had elevations on 6 of 1,428 nights (0.4%). One child in the youngest group accounted for 8 of all 31 elevations, none after a qualifying high. No other child contributed more than two.
Checks on the ketone readings and overnight records
Handheld meters have limited precision, and 21 of the 31 elevated blood ketone readings were close to the cutoff, at 1.1–1.4 mmol/L. Of the 10 readings at or above 1.5 mmol/L, only one followed a qualifying high. Urine strips provided a separate measure on nights without a qualifying high. Among mornings with a recorded strip, 19 of 26 with elevated blood ketones had urine ketones graded “small” or higher, compared with 25 of 2,451 at or below the blood cutoff. Even when sensor glucose never exceeded 250 mg/dL, 10 of 13 elevated mornings with a strip met the urine threshold.
A longer look back from the time of the ketone test identified one qualifying high that the overnight monitoring session did not show. Of all 50 recorded morning elevations, 40 matched a ketone meter reading near the end of the session, and 29 of those had sufficient glucose sensor readings stored in the pump for the preceding 12 hours. Two of those 29 tests followed a qualifying high. Of the 21 elevated mornings with sufficient readings for both analyses, one qualified in both windows, one qualified only in the 12-hour window and 19 qualified in neither. The longer record came from the same sensor.
Incomplete glucose records mattered most for the higher ketone readings: only 10 of 24 measured results at or above 1.5 mmol/L had sufficient overnight coverage for the main analysis. At the primary cutoff of 1.0 mmol/L, 19 elevated mornings lacked sufficient coverage. Even if all 19 had followed a qualifying high, at most 21 of the 50 measured elevations would qualify, keeping the classifications for the other 31 fixed. This bound excludes the 106 nights without a blood ketone result. The companion pump shut-off trial in people aged 15–45 (PSO3) yielded just three elevated mornings in 1,454 nights with sufficient sensor readings, too few for a useful estimate in older pump users.
What the later trials recorded
In DCLP3 and DCLP5, where a glucose high was a reason to test, we looked back from the first elevated ketone test in each cluster of positive readings, regardless of time of day. Of 27 clusters in DCLP3 and 21 in DCLP5, 20 and 12, respectively, had complete glucose sensor records for the preceding six hours. Most of those first tests followed a qualifying glucose high within four hours: 11 of 20 in DCLP3 and 9 of 12 in DCLP5. The other 9 of 20 in DCLP3 and 3 of 12 in DCLP5 had no such high in that window.
The 2-in-31 overlap applies to nights the PSO4 families monitored; the trial restricted bedtime glucose and discouraged system use during illness. A qualifying high was an opportunity to check: ketones at the time of the high could differ from ketones at breakfast. The later trials’ records cannot show what prompted a test or what it would have shown on an untested day. The overlap therefore cannot serve as a correction factor for another trial. An elevated reading alone does not diagnose diabetic ketoacidosis, and these comparisons cannot tell whether testing prevents it.
To interpret counts of days with elevated ketones, readers need to know which days families tested and why. The same ketone cutoff can capture different occasions under different testing schedules.
Methods
We used the Jaeb Center’s public trial data and protocols. The exploratory analysis covered 2,007 randomized home monitoring sessions in PSO3 and 3,606 in PSO4, with automatic pauses both enabled and disabled. We examined four glucose rules: waking glucose above 250 mg/dL, overnight sensor glucose above 250 mg/dL, overnight sensor glucose above 300 mg/dL and the later trials’ sensor rule. We examined ketone cutoffs of 0.6, 1.0 and 1.5 mmol/L. For the main comparison we used the sensor rule specified independently in the later protocols and counted mornings strictly above 1.0 mmol/L. Each observation was a monitoring session with a recorded morning blood ketone result. To estimate uncertainty in the fraction of elevated mornings that followed a qualifying high, we resampled the 82 children with replacement 50,000 times, keeping all nights for each sampled child. We took the 2.5th and 97.5th percentiles of the resulting fractions. Only two elevations followed qualifying highs, so the interval is approximate.
For each session we sorted the distinct sensor readings by time from activation to shutdown, counting each timestamp once. Sufficient coverage required 85–110% of the number of readings expected at five-minute intervals, no internal gap over 20 minutes and a reading within 15 minutes of shutdown. We timed each run above 300 mg/dL from its first to its last high reading. A reading at or below 300 mg/dL or a gap over 7.5 minutes ended a run. To find a confirming fingerstick, we searched the study glucose meter from 15 minutes before to 30 minutes after any qualifying reading, stopping no later than 15 minutes after shutdown. For the broader fingerstick count, we included any downloaded reading above 300 mg/dL from startup through 15 minutes after shutdown or a reported waking value above 300 mg/dL.
For the 12-hour check, we matched each child’s reported blood ketone concentration to the closest meter test in time with the same concentration within 30 minutes of shutdown. We excluded quality-control tests. We examined pump readings for the 12 hours ending at the matched test, applying the same coverage and run rules and requiring a reading within 15 minutes of each end of the window. We removed sensor calibration entries and converted metric glucose readings to mg/dL.
For DCLP3 and DCLP5 we used study-meter tests during randomized follow-up through the endpoint visit. We excluded quality-control tests and used corrected meter times when available. An elevated test began a cluster if it was the first or if more than 24 hours had passed since the previous elevated test. For each participant on each day in DCLP5, we used the sensor receiver with the most distinct valid readings. We required no gap over 7.5 minutes, including at the edges of the six hours before the test. A run above 300 mg/dL could start before the four-hour search window but had to meet the duration criterion within it, unless a single reading reached 400 mg/dL.