ReproductionImaging

Low-radiation imaging keeps added procedure time below five minutes overall and in four of six groups

Version 1 Published 2026-10-30 DOI 10.5281/zenodo.24373605

You are reading version 1. Read the latest version (2).

Original paper Generative AI-based low-dose digital subtraction angiography for intra-operative radiation dose reduction: a randomized controlled trial Zhao et al. (2026). Nature Medicine. doi:10.1038/s41591-025-04042-6
Numbers reproduced
Partly, error in the original
Original conclusion
Weakened, error in the original

The pooled time inference and subgroup confidence intervals reproduce, but the HCC noninferiority P value is incompatible with the stated direction and margin; only the procedure-specific time assurance is weakened, not the pooled conclusion.

In plain English

The original trial found that AI-assisted low-dose imaging reduced radiation and met a five-minute time limit when all procedures were combined. We checked that limit in each of six published procedure groups. Four met it; the lung- and liver-cancer treatment groups left larger delays possible, and the liver-cancer group’s favorable statistical result is inconsistent with its reported times. Neither group was proved to take more than five extra minutes on average. The overall result holds, but it offers less reassurance for these cancer treatments than the favorable liver-cancer result suggests.

GenDSA-V2 meets its trial-wide five-minute operation-time margin, but the published summaries establish the same margin in only four of six procedure subgroups. In the liver-cancer subgroup, AI increased mean time by 7.30 minutes (95% confidence interval 2.27–12.33); the correctly directed noninferiority P value is 0.816, incompatible with the reported P<0.001. The lung-cancer subgroup also leaves the time margin unresolved. These results narrow the procedure-specific reassurance available from a trial of radiation-reducing imaging, while preserving its pooled time conclusion. Neither cancer subgroup establishes a delay exceeding five minutes.

Background

Image-guided procedures require radiation, and reducing exposure is valuable if the workflow remains usable. Zhao and colleagues evaluated GenDSA-V2, a generative AI approach to low-dose digital subtraction angiography, in a randomized trial of 1,068 participants. Their abstract reports reduced radiation and mean operation times of 33.1 minutes with standard protocols and 34.8 minutes with AI, with P<0.001 for noninferiority. A randomized comparison supplies stronger evidence for clinical adoption than image-quality ratings alone. Its pooled estimate, however, averages across different procedures. [1]

Noninferiority asks whether an increase in time is smaller than a chosen acceptable margin. The authors specified five minutes and a one-sided significance threshold of 0.025 in Supplementary Appendix 2. To meet that criterion, the upper confidence bound for AI minus standard time must fall below five minutes. A mean difference below five is insufficient if its uncertainty still includes larger delays. Conversely, failure to meet the criterion does not establish that the delay exceeds five minutes. [2]

We rechecked the complete set of six indication/procedure groups in Supplementary Table 4: intracranial aneurysms (IA), intracranial atherosclerotic stenosis (ICAS), lung cancer (LC), hepatocellular carcinoma (HCC), percutaneous transthoracic needle biopsy (PTNB), and percutaneous liver biopsy (PLB). The first two groups concern cerebral angiography; the cancer interventions include bronchial and hepatic arterial chemoembolization, which delivers treatment through arteries. The last two groups concern needle biopsy. We retained the published grouping rather than selecting subgroups by their results. Operation time was a secondary endpoint. [2,3]

Results

Four subgroups establish the time margin

The cerebral angiography and biopsy groups meet the five-minute margin, including after adjustment across all six time comparisons. The cancer-treatment groups have wider intervals: upper bounds are 10.21 minutes for LC and 12.33 minutes for HCC (Figure 1). Their published mean radiation measures nevertheless fall by roughly two-thirds (Table 1). Those reductions describe a separate benefit; they do not resolve uncertainty about time.

Operation-time differences and confidence intervals by subgroup and overall

Figure 1. Mean operation-time differences, AI minus standard, with unadjusted two-sided 95% Welch confidence intervals. Positive differences mean longer procedures with AI. Each interval's upper endpoint is also a one-sided 97.5% upper confidence bound; noninferiority requires it to lie below the dashed five-minute line. The overall estimate uses the published overall summaries. Subgroup comparisons are exploratory.

Table 1. Reconstructed time contrasts and descriptive radiation reductions.

Subgroup Standard / AI n Standard / AI mean (SD), min Difference (95% CI), min NI P BH-adjusted P AK / DAP reduction
Cerebral aneurysm (IA) 89 / 101 34.1 (5.4) / 35.1 (5.9) +1.00 (-0.62, 2.62) 1.14e-06 3.41e-06 67.5% / 67.3%
Cerebral artery stenosis (ICAS) 129 / 116 34.8 (5.5) / 35.2 (5.8) +0.40 (-1.03, 1.83) 5.39e-10 3.24e-09 68.2% / 68.9%
Lung cancer (LC) 40 / 43 40.7 (16.0) / 44.2 (14.6) +3.50 (-3.21, 10.21) 0.329 0.394 67.5% / 68.6%
Liver cancer (HCC) 68 / 65 41.3 (14.7) / 48.6 (14.6) +7.30 (2.27, 12.33) 0.816 0.816 70.0% / 71.0%
Lung needle biopsy (PTNB) 126 / 131 27.1 (9.2) / 28.8 (9.3) +1.70 (-0.57, 3.97) 0.0023 0.00345 59.1% / 63.1%
Liver biopsy (PLB) 83 / 77 28.2 (9.0) / 27.0 (9.2) -1.20 (-4.05, 1.65) 1.47e-05 2.95e-05 61.4% / 64.1%

NI denotes the one-sided noninferiority test against a five-minute increase. BH denotes Benjamini–Hochberg adjustment across the six time tests. SD is standard deviation. Air kerma (AK) measures radiation energy delivered per unit mass of air; dose-area product (DAP) also accounts for the exposed area. Percentage reductions are calculated from the published arm means. They are not retests of radiation superiority or measurements of clinical harm. Counts and summaries come from Supplementary Tables 2 and 4. [2]

All four successful comparisons also survive Bonferroni adjustment, which controls the chance of any false-positive result across the six tests. The largest adjusted P among those four is 0.0138, below 0.025.

The favorable liver-cancer P value is wrong for the stated test

The HCC noninferiority result is inconsistent with the authors' own effect estimate and margin. Zhao and colleagues report a 7.3-minute increase, a 2.3–12.3-minute confidence interval, and an adjusted noninferiority P<0.001 in Supplementary Table 4. Its footnote explicitly identifies operation-time comparisons as noninferiority tests. [2] Because the estimated increase already exceeds five minutes, the correctly directed test has P>0.5. Multiplicity adjustment cannot turn that into P<0.001. Our reconstructed P is 0.816.

We therefore judge this reported favorable subgroup P value to be an error, rather than a consequence of unavailable participant records. The public summaries cannot distinguish an incorrect tail or sign from a reporting mistake. The HCC interval supports a positive mean delay, but crosses the five-minute boundary; testing for a delay greater than five minutes gives P=0.184. The corresponding LC test gives P=0.671.

The reported subgroup differences and intervals largely reproduce. Normal-approximation intervals match all six published intervals to their displayed precision. PTNB's reported P<0.001 also differs from our unadjusted P=0.00230, although the subgroup still meets the margin after adjustment. LC's reported adjusted P=0.981 differs from our BH value of 0.394. The original adjustment family is unclear, so that difference alone does not identify an error.

Pooled noninferiority remains supported

Using the published overall means and standard deviations gives a 1.70-minute increase (95% CI 0.34–3.06), with noninferiority P=1.07×10⁻⁶. Independently combining the six subgroup summaries recovers all 535 standard-arm and 533 AI-arm participants and gives a similar 1.67-minute increase (95% CI 0.32–3.02). The small discrepancy between reconstructed and published overall AI standard deviations is compatible with rounding of the subgroup inputs.

Combining LC and HCC gives a 5.77-minute increase (95% CI 1.77–9.77). The original supplement already reports a nonsignificant time noninferiority result for their broader two-dimensional thorax/abdomen category in Table 18. [2] Thus, the correction concerns the favorable HCC result and the complete procedure-level accounting; it does not overturn the trial-wide estimate or newly reveal the broader category's uncertainty.

The subgroup decisions remain unchanged with equal-variance t tests, normal approximations, and perturbations of the displayed means and standard deviations by half their last reported digit. The four successful comparisons remain Bonferroni-significant at every evaluated rounding corner.

Interpretation

The trial supports a less-than-five-minute mean time increase overall and within its cerebral angiography and biopsy subgroups. For the two cancer-treatment subgroups, the summaries allow delays both below and above that boundary. The published favorable HCC P value overstates the available time evidence. This is a measured correction to procedure-specific inference, with the pooled conclusion intact.

These exploratory subgroups were not six separately powered trials. We assume the baseline arm counts are the time-analysis denominators and treat participants as independent. Public summaries cannot resolve missing outcomes, center effects, skewed time distributions, covariate adjustment, or agreement between intention-to-treat and per-protocol analyses. Differences in subgroup P values do not establish a treatment-by-procedure interaction. The analysis concerns average operation time, not individual delays or clinical safety.

Methods

We used anonymously downloadable Supplementary Information and the Reporting Summary. Arm sizes were transcribed from Supplementary Table 2; time means, standard deviations and radiation means came from Table 4. Both tables were checked against rendered PDF pages. All six groups were fixed before calculation. The three source-data workbooks contain image ratings and other figure data rather than individual randomized operation times. [2,3]

For each subgroup, we subtracted the standard-arm mean from the AI-arm mean. The standard error was the square root of the sum of the two arm variances divided by their respective sample sizes. We used Welch–Satterthwaite degrees of freedom. The noninferiority statistic was the mean difference minus five minutes, divided by its standard error; its lower-tail t probability tests the null that AI adds at least five minutes. We used a one-sided threshold of 0.025 and two-sided 95% confidence intervals. A separate upper-tail probability tests whether the increase exceeds five minutes.

The supplement states false-discovery-rate adjustment and identifies Benjamini–Hochberg in Table 18, but does not unambiguously define Table 4's adjustment family. We applied BH to the six operation-time tests and compared adjusted P values with 0.025. Bonferroni adjustment over those same six tests provided a stricter sensitivity analysis. Neither calculation is claimed to reconstruct the authors' unknown adjustment family exactly.

Sensitivity analyses used equal-variance t and normal tests. We enumerated all corner combinations of means and standard deviations shifted by ±0.05 minutes; LC's standard-arm SD, printed as 16, was also varied by ±0.5. Overall moments were reconstructed using sample-size-weighted means and both within- and between-subgroup sums of squares. Rounding checks for pooled moments enumerated all 4,096 mean/SD corner combinations per arm at ±0.05. Radiation reductions were one minus the AI-to-standard ratio of published means. Calculations used Python 3.12.16, NumPy 2.5.4 and SciPy 1.18.1.

Appendix 2 provides the margin and significance level, but we could not independently verify their prospective specification. The Reporting Summary says a protocol was submitted to the journal; we found no separately downloadable protocol, and direct registry retrieval failed. [3,4] A repeat article-history and Crossmark check on 31 October 2026 UTC found an affiliation/address update dated 29 January 2026. Targeted searches found no time-related correction or letter. [1,5]

References

  1. Zhao H, Bai Y, Chen L, et al. Generative AI-based low-dose digital subtraction angiography for intra-operative radiation dose reduction: a randomized controlled trial. Nature Medicine. 2026;32:288–296. Article.
  2. Zhao H, et al. Supplementary Information, Appendix 2 and Tables 2, 4 and 18.
  3. Zhao H, et al. Reporting Summary.
  4. Chinese Clinical Trial Registry. ChiCTR2400084789.
  5. Crossref. Crossmark update record.