Low-radiation imaging keeps added procedure time below five minutes overall and in four of six groups
- Numbers reproduced
- Partly, error in the original
- Original conclusion
- Weakened, error in the original
The pooled time inference and subgroup confidence intervals reproduce, but the HCC noninferiority P value is incompatible with the stated direction and margin; only the procedure-specific time assurance is weakened, not the pooled conclusion.
In plain English
The original trial found that AI-assisted low-dose imaging reduced radiation and met a five-minute time limit when all procedures were combined. We checked that limit in each of six published procedure groups. Four met it; the lung- and liver-cancer treatment groups left larger delays possible, and the liver-cancer group’s favorable statistical result is inconsistent with its reported times. Neither group was proved to take more than five extra minutes on average. The overall result holds, but it offers less reassurance for these cancer treatments than the favorable liver-cancer result suggests.
Reducing radiation during image-guided procedures is valuable, but clinicians also need to know whether a lower-dose approach makes procedures take longer. A trial-wide average can hide differences between procedures. We rechecked a randomized trial of GenDSA-V2, an artificial intelligence (AI) approach to low-dose imaging. The published summaries support an average increase below the trial's five-minute limit overall and in four of six procedure groups. For liver-cancer treatment, the estimated increase was 7.30 minutes (95% confidence interval 2.27–12.33). Its correctly directed noninferiority test, which asks whether the increase is below the acceptable limit, gives P=0.816, incompatible with the reported P<0.001. The time limit remains unresolved for both cancer-treatment groups: their results allow delays below or above five minutes.
Background
Image-guided procedures require radiation, and reducing exposure is valuable if the workflow remains usable. Zhao and colleagues evaluated GenDSA-V2 in a randomized trial of 1,068 participants. The system uses generative AI for low-dose digital subtraction angiography, an X-ray technique that subtracts background structures to make blood vessels visible. Their abstract reports reduced radiation and mean operation times of 33.1 minutes with standard protocols and 34.8 minutes with AI, with P<0.001 for noninferiority. A randomized comparison supplies stronger evidence for clinical adoption than image-quality ratings alone. Its pooled estimate, however, averages across different procedures. [1]
Noninferiority asks whether an increase in time is smaller than a chosen acceptable margin. The authors specified five minutes and a one-sided significance threshold of 0.025 in Supplementary Appendix 2. To meet that criterion, the upper confidence bound for AI minus standard time must fall below five minutes. A mean difference below five is insufficient if its uncertainty still includes larger delays. [2]
We rechecked all six indication/procedure groups in Supplementary Table 4. Intracranial aneurysms (IA) are bulges in brain arteries; intracranial atherosclerotic stenosis (ICAS) is narrowing of those arteries by plaque. Both groups underwent cerebral angiography, which images blood vessels in the brain. The lung cancer (LC) and hepatocellular carcinoma (HCC, a type of liver cancer) interventions include bronchial and hepatic arterial chemoembolization. These procedures deliver cancer treatment through arteries and block blood supply to the treated tissue. Percutaneous transthoracic needle biopsy (PTNB) and percutaneous liver biopsy (PLB) take tissue samples through the skin from the lung and liver. We retained all published groups before calculating results. Operation time was a secondary outcome of the trial. [2,3]
Results
Four subgroups establish the time margin
The cerebral angiography and biopsy groups meet the five-minute margin, including after adjustment across all six time comparisons. The cancer-treatment groups have wider intervals: upper bounds are 10.21 minutes for LC and 12.33 minutes for HCC (Figure 1). Their published mean radiation measures nevertheless fall by roughly two-thirds (Table 1). Those reductions describe a separate benefit.

Figure 1. Mean operation-time differences, AI minus standard, with unadjusted two-sided 95% Welch confidence intervals, which allow the two arms to have different variances. Positive differences mean longer procedures with AI. Each interval's upper endpoint is also a one-sided 97.5% upper confidence bound; noninferiority requires it to lie below the dashed five-minute line. The overall estimate uses the published overall summaries. Subgroup comparisons are exploratory.
Table 1. Reconstructed time differences and descriptive radiation reductions. Sample size is denoted by n, standard deviation by SD, and confidence interval by CI. NI denotes the one-sided noninferiority test against a five-minute increase. BH denotes Benjamini–Hochberg adjustment across the six time tests, which controls the false discovery rate: the expected proportion of false-positive results among tests declared significant. Air kerma (AK) measures radiation energy delivered per unit mass of air; dose-area product (DAP) also accounts for the exposed area.
| Subgroup | Standard / AI n | Standard / AI mean (SD), min | Difference (95% CI), min | NI P | BH-adjusted P | AK / DAP reduction |
|---|---|---|---|---|---|---|
| Cerebral aneurysm (IA) | 89 / 101 | 34.1 (5.4) / 35.1 (5.9) | +1.00 (-0.62, 2.62) | 1.14e-06 | 3.41e-06 | 67.5% / 67.3% |
| Cerebral artery stenosis (ICAS) | 129 / 116 | 34.8 (5.5) / 35.2 (5.8) | +0.40 (-1.03, 1.83) | 5.39e-10 | 3.24e-09 | 68.2% / 68.9% |
| Lung cancer (LC) | 40 / 43 | 40.7 (16.0) / 44.2 (14.6) | +3.50 (-3.21, 10.21) | 0.329 | 0.394 | 67.5% / 68.6% |
| Liver cancer (HCC) | 68 / 65 | 41.3 (14.7) / 48.6 (14.6) | +7.30 (2.27, 12.33) | 0.816 | 0.816 | 70.0% / 71.0% |
| Lung needle biopsy (PTNB) | 126 / 131 | 27.1 (9.2) / 28.8 (9.3) | +1.70 (-0.57, 3.97) | 0.0023 | 0.00345 | 59.1% / 63.1% |
| Liver biopsy (PLB) | 83 / 77 | 28.2 (9.0) / 27.0 (9.2) | -1.20 (-4.05, 1.65) | 1.47e-05 | 2.95e-05 | 61.4% / 64.1% |
Percentage radiation reductions are calculated from the published arm means. Counts and summaries come from Supplementary Tables 2 and 4. [2]
All four successful comparisons also survive Bonferroni adjustment, which controls the chance of any false-positive result across the six tests. The largest adjusted P among those four is 0.0138, below 0.025.
The favorable liver-cancer P value is wrong for the stated test
The HCC noninferiority result is inconsistent with the authors' own effect estimate and margin. Zhao and colleagues report a 7.3-minute increase, a 2.3–12.3-minute confidence interval, and an adjusted noninferiority P<0.001 in Supplementary Table 4. Its footnote explicitly identifies operation-time comparisons as noninferiority tests. [2] Because the estimated increase already exceeds five minutes, the correctly directed test has P>0.5. Adjustment for multiple tests cannot turn that into P<0.001. Our reconstructed P is 0.816.
The reported favorable subgroup P value is therefore an error for the stated test. The HCC interval supports a positive mean delay, but crosses the five-minute boundary; testing for a delay greater than five minutes gives P=0.184. The corresponding LC test gives P=0.671.
The reported subgroup differences and intervals largely reproduce. Normal-approximation intervals match all six published intervals to their displayed precision. PTNB's reported P<0.001 also differs from our unadjusted P=0.00230, although the subgroup still meets the margin after adjustment. LC's reported adjusted P=0.981 differs from our BH value of 0.394. The set of comparisons included in the original adjustment is unclear, so that difference alone does not identify an error.
The trial-wide result remains supported
Using the published overall means and standard deviations gives a 1.70-minute increase (95% CI 0.34–3.06), with noninferiority P=1.07×10⁻⁶. Independently combining the six subgroup summaries recovers all 535 standard-arm and 533 AI-arm participants and gives a similar 1.67-minute increase (95% CI 0.32–3.02). The small discrepancy between reconstructed and published overall AI standard deviations is compatible with rounding of the subgroup inputs.
Combining LC and HCC gives a 5.77-minute increase (95% CI 1.77–9.77). The original supplement already reports a nonsignificant time noninferiority result for their broader two-dimensional thorax/abdomen (chest/belly) imaging category in Table 18. [2] The correction concerns the favorable HCC result and the accounting for all six procedure groups; the trial-wide estimate remains supported.
The subgroup decisions remain unchanged with equal-variance t tests, normal approximations, and perturbations of the displayed means and standard deviations by half their last reported digit. The four successful comparisons remain significant after Bonferroni adjustment at every evaluated combination of rounding limits.
Methods
We used the publicly downloadable Supplementary Information and Reporting Summary, which describe the trial methods and results. Arm sizes were transcribed from Supplementary Table 2; time means, standard deviations and radiation means came from Table 4. Both tables were checked visually against the original PDF pages. All six groups were fixed before calculation. The three source-data workbooks contain image ratings and other figure data rather than individual randomized operation times. [2,3]
For each subgroup, we subtracted the standard-arm mean from the AI-arm mean. The standard error was the square root of the sum of the two arm variances divided by their respective sample sizes. We used the Welch–Satterthwaite approximation to obtain the degrees of freedom for the t distribution, allowing unequal arm variances. The noninferiority statistic was the mean difference minus five minutes, divided by its standard error; its lower-tail t probability tests the null that AI adds at least five minutes. We used a one-sided threshold of 0.025 and two-sided 95% confidence intervals. A separate upper-tail probability tests whether the increase exceeds five minutes.
The supplement states false-discovery-rate adjustment and identifies Benjamini–Hochberg in Table 18, but does not unambiguously define which comparisons were included in Table 4's adjustment. We applied BH to the six operation-time tests and compared adjusted P values with 0.025. Bonferroni adjustment over those same six tests provided a stricter sensitivity analysis.
Sensitivity analyses used equal-variance t and normal tests. We evaluated every combination of upper and lower rounding limits for means and standard deviations, shifting each by ±0.05 minutes; LC's standard-arm SD, printed as 16, was also varied by ±0.5. Overall means and variances were reconstructed using sample-size-weighted means and sums of squared deviations both within and between subgroups. Rounding checks for these combined summaries evaluated all 4,096 combinations of upper and lower mean/SD rounding limits per arm at ±0.05. Radiation reductions were one minus the AI-to-standard ratio of published means. Calculations used Python 3.12.16, NumPy 2.5.4 and SciPy 1.18.1.
Appendix 2 provides the margin and significance level. The Reporting Summary says a protocol was submitted to the journal; we found no separately downloadable protocol, and direct retrieval from the Chinese Clinical Trial Registry failed. [3,4] An article-history check and a check of Crossmark, a service that records publication updates, on 31 October 2026 UTC found an affiliation/address update dated 29 January 2026. Targeted searches found no time-related correction or letter. [1,5]
Interpretation
The trial supports a less-than-five-minute mean time increase overall and within its cerebral angiography and biopsy groups. The published favorable HCC P value overstates the procedure-specific time evidence. The radiation reductions remain a separate benefit, while the time results leave uncertainty about adopting the workflow for cancer treatment.
These exploratory subgroups were not six separately powered trials. We assume baseline arm counts are the time-analysis denominators and treat participants as independent. Public summaries cannot resolve missing outcomes, differences between centers, skewed time distributions, adjustment for participant characteristics, or agreement between analyses of all randomized participants and those who followed the protocol. They also cannot distinguish a wrong test direction from a reporting mistake in HCC. Prospective specification of the margin and threshold could not be independently verified, and the authors' exact multiple-testing adjustment remains unknown. Different subgroup P values alone do not establish that the treatment effect differs by procedure. The results concern average time, not individual delays or clinical safety; the radiation percentages are descriptive reductions, not new statistical tests of radiation benefit or measurements of harm.
References
- Zhao H, Bai Y, Chen L, et al. Generative AI-based low-dose digital subtraction angiography for intra-operative radiation dose reduction: a randomized controlled trial. Nature Medicine. 2026;32:288–296. Article.
- Zhao H, et al. Supplementary Information, Appendix 2 and Tables 2, 4 and 18.
- Zhao H, et al. Reporting Summary.
- Chinese Clinical Trial Registry. ChiCTR2400084789.
- Crossref. Crossmark update record.