Combining scores from two wrist X-ray views added little fracture detection beyond the side view
You are reading version 1. Read the latest version (2).
- Numbers reproduced
- No, limits of our recheck
- Original conclusion
- Unclear
The released checkpoint could not be linked to exact training patients, so we evaluated a fresh single-class model; the original numerical improvement and clinical claims were not adjudicated.
In plain English
The original study reported improved AI detection on individual wrist X-rays. We trained a fresh detector and checked whether taking the higher score from two views helped it find more fracture-positive examinations without adding false alarms. The two-view rule added a small, uncertain benefit over the lateral view selected using separate development data. This does not show that children need fewer X-rays: the study tested one AI rule against public annotations, not clinical care.
Combining two X-ray views gave a small, uncertain improvement over a lateral-view fracture detector in a public pediatric wrist dataset. Among 485 held-out examinations, taking the higher of the two image scores detected three additional fracture-positive examinations but produced three additional false alarms. At a common 95% specificity, the sensitivity gain was 0.9 percentage points (95% patient-bootstrap interval −1.7 to +3.5). The subgroup with a fracture annotation in only one view showed no clear gain. This fresh-model extension shows why evaluating a second view requires counting both recovered fractures and false alarms at the examination level.
Background
A wrist examination contains different projections of the same anatomy. Here, PA/AP denotes the front-facing projection, acquired with the beam passing from back to front or front to back; the lateral projection is the side view. A detector may assign a high fracture score to one image and a low score to the other. Taking the higher score, called max-score fusion, is a simple way to combine them. It can recover positive examinations missed by one projection, but it also gives each negative examination two opportunities to trigger an alert.
Ju and Cai reported improved object detection on GRAZPEDWRI-DX using YOLOv8 and image augmentation. Their headline mean average precision at an overlap threshold of 0.5 was 0.638, compared with 0.636 for the original YOLOv8 and 0.634 for an improved YOLOv7. This metric summarizes detection precision and recall across annotated object classes and score thresholds. Their public data partitions and detector release provide a basis for subsequent checks, but the image-level metric does not directly answer how often an entire examination is missed or falsely flagged. [1]
We asked whether max-score fusion improves examination-level fracture detection over a single projection selected using separate development data. Sensitivity is the fraction of reference-positive examinations flagged; specificity is the fraction of reference-negative examinations left unflagged. We gave particular attention to examinations with a fracture box in only one projection, where complementary evidence might be most useful. A box is an annotator's rectangle around a fracture, not an independent measure of everything visible in the image.
We could not verify which patients trained the released wrist checkpoint, so we trained a smaller, single-class detector on documented patient-disjoint partitions. We do not conclude that Ju and Cai's published numbers are wrong. The numerical reproduction remains unresolved because of this provenance limit and our different model and training population. The result below concerns the fresh detector and the simple two-score combination.
Results
The paired test set included 44 positives annotated in one view
The primary analysis contained 4,711 examinations with one PA/AP and one lateral image, both marked as initial examinations and without positive cast or metal tags. The released partitions supplied 3,278 training, 948 development and 485 test examinations; no patient crossed partitions. We excluded 245 examinations carrying an uncertain-diagnosis tag before fitting or primary evaluation.
The test set contained 438 patients, 228 reference-positive examinations and 257 reference-negative examinations. A positive reference meant a fracture box or an AO fracture-classification code on either image. Among the positives, 182 had boxes in both views, 44 in one view and two had AO codes without boxes. Of the 44 one-view positives, 39 had lateral-only boxes and five had PA/AP-only boxes. Thus, the opportunity for a front-facing image to add an explicitly annotated fracture to a lateral-only detector was limited in this cohort.
Frozen thresholds yielded three additional detections and three additional false alarms
Lateral was the better single projection on development data and was selected before test evaluation. At thresholds targeting 95% development specificity, lateral detected 425 of 453 positive examinations, compared with 393 for PA/AP. Each policy, including fusion, falsely flagged 24 of 495 development negatives, giving 95.15% specificity.
Table 1 shows the policies applied unchanged to test examinations. Fusion increased sensitivity from 90.4% to 91.7%, a difference of 1.3 percentage points (95% interval −0.9 to +3.2). Specificity fell from 94.6% to 93.4%, a difference of −1.2 points (−2.9 to +2.8). Equal development targets therefore did not produce equal test false-alarm rates.
Table 1. Test performance at separately calibrated, frozen thresholds targeting 95% development specificity. The same examinations enter every row.
| Policy | Positives detected / 228 | Sensitivity | False alarms / 257 | Specificity | One-view positives detected / 44 |
|---|---|---|---|---|---|
| PA/AP alone | 181 | 79.4% | 13 | 94.9% | 12 |
| Lateral alone (development-selected) | 206 | 90.4% | 14 | 94.6% | 32 |
| Maximum of two views | 209 | 91.7% | 17 | 93.4% | 33 |
Fusion rescued four positive examinations that lateral missed and lost one that lateral flagged. Three rescues had an above-threshold prediction overlapping a reference fracture box. One was a one-view-annotated examination, rescued by a localized PA/AP detection; the other three had boxes in both views. The lost examination had an AO code but no fracture box. Lateral flagged both AO-only test positives, whereas fusion flagged one. Although a maximum score cannot fall below either component score, fusion's higher calibrated threshold can lose a detection. The AO-only flags provide no evidence of correct localization.
The gain remained small after matching specificity
At common 95% test specificity, fusion sensitivity was 90.8% and lateral sensitivity was 89.9% (Table 2). The difference was 0.9 percentage points (95% interval −1.7 to +3.5). These descriptive estimates interpolate between neighboring test score thresholds. They answer the equal-false-alarm question but are not thresholds that could have been chosen without test labels.
Table 2. Sensitivity at common 95% test specificity. The one-view subgroup uses the same operating points as the full cohort, calibrated against all 257 test negatives.
| Positive examinations | Lateral sensitivity | Two-view sensitivity | Difference, percentage points (95% interval) |
|---|---|---|---|
| All 228 | 89.9% | 90.8% | +0.9 (-1.7 to +3.5) |
| 44 annotated in one view | 72.7% | 70.5% | -2.3 (-8.9 to +7.3) |
The one-view subgroup's point estimate changed direction after matching specificity (Figure 1). Frozen thresholds detected 33 of 44 with fusion and 32 with lateral. At common specificity, sensitivity was 70.5% versus 72.7%, a difference of −2.3 points (−8.9 to +7.3). This interval leaves both benefit and harm plausible; it does not establish equivalence.

Figure 1. Two-view minus lateral sensitivity, with 95% patient-bootstrap intervals. Positive values favor two-view fusion. Frozen thresholds target 95% specificity on development data but achieve different test specificities. Common-specificity estimates use test ROC interpolation. Each bootstrap independently resamples development and test patients; all repeat examinations remain together. Lateral was selected in all 2,000 replicates.
The secondary analyses also gave small net sensitivity differences. At the 90% development target, fusion detected 212 of 228 positives versus 211 for lateral, with 23 versus 29 false alarms. The common-90%-specificity sensitivity difference was +0.4 points (−1.3 to +3.9). Across the prespecified 90–99% common-specificity grid, the overall contrast ranged from −2.6 to +3.1 points. Including the 22 uncertain test examinations, without changing primary thresholds, gave sensitivity of 90.6% versus 89.3% and specificity of 93.0% versus 94.1%, with the same four rescues and one loss.
Discussion
For this detector, the lateral projection already captured much of the annotation-defined fracture signal. Adding the front-facing score produced a small net gain, and matching the false-alarm burden did not reveal a clear advantage in the subgroup expected to benefit from complementary views. Comparing fusion only with PA/AP would have given a much larger apparent benefit than comparing it with the development-selected projection. The practical evaluation lesson is to choose the single-view comparator without test truth and to measure the achieved examination-level false-alarm rate.
The evidence is limited to one freshly trained detector, one hospital dataset and a simple maximum-score rule. The 44-examination subgroup remains too small to exclude a clinically useful gain. Absence of a fracture annotation does not establish clinical absence of fracture, and a box-free projection can still contain diagnostic information. The uncertainty intervals condition on one trained model, excluding retraining variability. Larger detectors or learned multiview combinations may behave differently. These results cannot justify dropping a clinical projection or guide discharge decisions.
Methods
Data and reference labels
We used GRAZPEDWRI-DX version 2, its image metadata and Pascal VOC fracture boxes. The dataset contains 20,327 pediatric wrist radiographs collected at the University Hospital Graz. Its description distinguishes known fractures from fractures localized on a particular image. [2,3]
We grouped images by patient, study number and side. Of 10,842 groups, 5,761 had all images tagged initial, 5,752 also lacked positive cast/metal tags, and 4,956 contained exactly the required two projections. We also checked for metal object annotations; none occurred among eligible pairs. Projection timestamps could differ because the images were acquired sequentially. Complete treatment histories were unavailable. Excluding any pair with an uncertain-diagnosis tag gave the primary cohort. The original released patient partitions were retained. [4]
Reference-positive status was the union of fracture boxes and AO codes across both images. Across all 4,956 eligible pairs, 18 AO-coded examinations had no fracture boxes, while 29 box-positive examinations lacked AO codes. The primary test set contained two of the AO-only examinations. Reference-negative status meant neither annotation was present. The released fracture-visibility flag exactly matched box presence across all images, so it supplied no independent confirmation. For detector training, images without fracture boxes received empty detection labels even if their examination carried an AO code.
Detector training and inference
The released wrist checkpoint contained a March 2023 training date, while the current partition files first appeared publicly in October 2023. No manifest linked its training patients to those partitions. This chronology leaves linkage unresolved; it is not evidence of patient leakage. We instead initialized YOLOv8n with official COCO general-image weights and trained one fracture class on 6,556 images from primary training pairs. Development used 1,896 images.
Preprocessing followed the released conversion notebook: divide 16-bit intensities by 65,535, rescale using the 0.05th and 99.95th percentiles, apply contrast-limited adaptive histogram equalization with scikit-image defaults, then normalize to 8-bit intensities. Original geometry was preserved before model resizing. All 4,595 training and development boxes agreed with the alternative YOLO annotation export to within one pixel at matched edges.
We used Ultralytics 8.3.40, PyTorch 2.6.0 and NumPy 1.26.4. Training used 640-pixel inputs, batch size eight, AdamW, initial learning rate 0.001, final learning-rate multiplier 0.01, weight decay 0.0005 and zero warmup bias learning rate. Augmentation allowed rotations up to 10 degrees, translation fraction 0.1, scale variation 0.2 and horizontal flips with probability 0.5. Vertical flips, color augmentation, mosaic and mixup were disabled. The seed was 20261030. The maximum was 30 epochs with early-stopping patience ten; all 30 completed.
Checkpoint selection used development detection fitness: 0.1 times mean average precision at box intersection-over-union 0.5 plus 0.9 times the average over overlap thresholds 0.5–0.95. Intersection-over-union is the shared area of two boxes divided by their combined area. Epoch 29 was selected, with these development precision metrics of 0.93636 and 0.56395, respectively. These single-class development metrics are not comparable to the original multi-class headline score. The weights were locked before evaluation scoring.
Inference retained predictions with confidence at least 0.001, suppressed overlapping predictions at intersection-over-union 0.7, and retained at most 300 boxes per image. An image's score was the maximum fracture confidence, or zero if no prediction remained. Fusion used the higher image score. Reference boxes affected localization checks only. A localized rescue required an above-threshold prediction with intersection-over-union at least 0.5 against a fracture box in the corresponding image. Saved image scores and localized scores were independently recomputed from the full predictions.
Calibration and uncertainty
For each policy, we selected the lowest score threshold meeting at least the target specificity on development negatives, excluding entire tied score groups where necessary. Scores at or above the threshold triggered an alert. The single projection with greater development sensitivity was chosen, with ties assigned to PA/AP. Primary thresholds were approximately 0.119446 for PA/AP, 0.171256 for lateral and 0.250732 for fusion. Development data served both checkpoint selection and threshold calibration.
We evaluated paired sensitivity and specificity differences and counted rescues and losses on identical test examinations. For common-specificity comparisons, we linearly interpolated between adjacent empirical receiver operating characteristic (ROC) points, retaining both ends of vertical segments and choosing the highest sensitivity at an exact observed false-positive rate. Interpolated points represent randomized mixtures of neighboring threshold policies. Subgroup sensitivity was evaluated at those same endpoints, rather than fitting a separate subgroup operating point.
We independently resampled development and test patients 2,000 times with seed 20261030. Both views and repeat examinations remained within their patient cluster. Each replicate reselected the single projection, recalibrated thresholds and recomputed test ROC points. Reported intervals are the 2.5th and 97.5th bootstrap percentiles. The 90% target and the 90–99% specificity grid were prespecified. The uncertain-label inclusion analysis retained primary thresholds and reports point estimates only.
References
- Ju R-Y, Cai W. Fracture detection in pediatric wrist trauma X-ray images using YOLOv8 algorithm. Scientific Reports. 2023;13:20077. Article.
- Nagy E, et al. A pediatric wrist trauma X-ray dataset (GRAZPEDWRI-DX) for machine learning. Scientific Data. 2022;9:222. Article.
- GRAZPEDWRI-DX, version 2. Public dataset.
- Ju R-Y, Cai W. Bone Fracture Detection YOLOv8. Public detector repository and partitions.